GastroCH4ECK® hydrogen and methane breath test monitor

Portable breath methane, hydrogen and oxygen monitor to aid in the detection of gastrointestinal disorders.

What is it?

Helping detect gastrointestinal disorders, the GastroCH4ECK® is the newest member of the Gastrolyzer® family. It is a portable breath analysis monitor that measures accurate and real-time combined CH4, H2 and O2. This monitor is intended for multi-patient use by healthcare professionals in a clinical environment.

GastroCH4ECK® hydrogen and methane breath test monitor

Why is it important?

The monitor provides accurate and real-time combined CH4, H2 and O2 monitoring.

The GastroCH4ECK® is a portable breath CH4, H2 and Omonitor that helps health professionals to accurately detect a range of gastrointestinal disorders. An O2 reading is taken to motivate patients into providing an end-tidal sample. If they are unable to do this the GastroCH4ECK® will automatically correct the reading saving the patient from embarrassment and having to carry out another test.

A percentage of patients do not produce hydrogen and therefore would not be diagnosed by the use of hydrogen breath testing (HBT) alone. Therefore the role of methane is to ensure that patients who are non-hydrogen producers, but produce methane instead, are not miss-diagnosed.

Most monitors measure hydrogen alone, which identifies less than 50% of sufferers2 . Methane is detected in 30%-50% of the healthy adult population worldwide3

Why GastroCH4ECK?

The GastroCH4ECK® can be used to help detect the following disorders:

Lactose intolerance 
Fructose intolerance 
Glucose intolerance
Irritable Bowel Syndrome (IBS) 
Small Intestinal Bacterial Overgrowth (SIBO)

"If you are serious about providing a Gold Standard breath testing service, the Bedfont® GastroCH4ECK® system is essential, as without methane monitoring you will be missing very important clinical information and increase your percentage of false negative tests. It is also very important to map patients symptoms prior to and during the test to make sure the clinical translation of the physiological findings are maximised."

Dr Anthony Hobson from the Functional Gut Clinic

Benefits for the user

The monitor provides hydrogen and methane breath sampling with instant results.

The breath bag sampling options allows for simultaneous testing of larger groups or remote testing for patients who cannot provide samples directly in person.

The monitor only requires calibration once per month, saving precious time and calibration gas costs.

The monitor has a fast warm up time of less than 30 minutes.

The monitor provides O2 correction for accurate results.

There is a real-time traffic light dial and smiley face system to motivate patients and indicate how the test is going.

The consumables contain anti-bacterial filters for optimum infection control.

Bedfont Gastrolyzer® Range of Breath Monitors

A video introduction to our hydrogen and methane breath test monitors, to aid in the detection of gastrointestinal disorders such as lactose intolerance, small intestinal bacterial overgrowth (SIBO) and irritable bowel syndrome (IBS).

 

How to use Gastrolyzer breath bags

This is a simple illustrative demonstration of how to use the Gastrolyzer® breath bags.

GastroCH4ECK® Consumables

GastroCH4ECK® Mouthpieces

The GastroCH4ECK® breath sample mouthpiece enables patients to exhale directly into the monitor to get instant results. The mouthpiece itself is specifically designed with the latest bacterial filtration to remove 99.9% of airborne bacteria and any moisture from the patient’s breath. Mouthpieces are single patient use only.

Moisture Removal Filters

For use with the breath bag sampling mode. These will last for approximatley 150 samples and need to be changed when the filter changes from orange to dark green.

GastroCH4ECK® Breath Bags

Specifically designed to capture as much end-tidal air as possible, with a one-way mouthpiece to prevent loss of sample. GastroCH4ECK® breath bags are 'one patient per use' so can be used for the duration of a patient's testing procedure and then disposed of.

Monitor cleaning wipes

Free from alcohol, ensuring the continued performance of your monitor. Pack contains 50 wipes.

GastroCH4ECK® Calibration Kit

GastroCH4ECK® requires calibration once per month using 100ppm H2, 100ppm CH4 & 20.9% air provided as a kit or replacement cylinder.

Product specification

Power supply
Mains power 230v [110v compatible]
Warm-up time
30 minutes
Calibration frequency
Once a month
Detection principle
Electrochemical and Optical sensors
Operating temperature
15 - 35 degrees oC
Operating humidity
30 – 75% non-condensing
Dimensions
300 x 265 x 140 mm
Weight
Approx. 6kg
Gases measured
CH4 (Methane) H2 (Hydrogen) O2 (Oxygen)
Range
CH4 0 – 200 ppm, H2 0 – 200 ppm, O2 0 – 100%
Accuracy: CH4 resolution
1 ppm Accuracy: +/- (10% of range or 10% of reading)*H2
Resolution: 1 ppm accuracy
± 10%**O2 Resolution: 0.1% Accuracy: +/-2% full scale
Test time
CH4/H2/O2 typically < 45 seconds

Testimonials

"If you are serious about providing a gold-standard breath testing service, the Bedfont® GastroCH4ECK® system is essential, as without methane monitoring you will be missing very important clinical information and increase your percentage of false negative tests. It is also very important to map patients symptoms prior to and during the test to make sure the clinical translation of the physiological findings are maximised."
- Dr Hobson, The Functional Gut Clinic

Downloads

Please click on the links to the right to view product information and manuals.

Join the Gastrolyzer forum

Did you know purchasing this product entitles you to free membership of the Gastrolyzer® Email Forum.
 
What is the Gastrolyzer® email forum?
It is an international, invitation only forum where professionals using the Bedfont Gastrolyzer® range of monitors can communicate and share knowledge.
There is no cost or obligation to participate but membership is free when you purchase a Bedfont Gastrolyzer® Gastro+TM or GastroCH4ECK® - so why not take advantage and join this elite group today!

For more information, please contact Don@bedfont.com. 

Clinical Papers

References

1. Reference G. lactose intolerance [Internet]. Genetics Home Reference. 2017 [cited 11 January 2017]. Available from: https://ghr.nlm.nih.gov/condition/lactose-intolerance#synonyms
2. Waud J, Matthews S, Campbell A. Measurement of breath hydrogen and methane, together with lactase genotype, defines the current best practice for investigation of lactose sensitivity. Annals of Clinical Biochemistry. 2008;45(1):50-58.
3. Triantafyllou K, Chang C, Pimentel M. Methanogens, Methane and Gastrointestinal Motility. Journal of Neurogastroenterology and Motility. 2014;20(1):31-40.
4. Pereira J, Porto-Figueira P, Cavaco C, Taunk K, Rapole S, Dhakne R et al. Breath Analysis as a Potential and Non-Invasive Frontier in Disease Diagnosis: An Overview. Metabolites. 2015;5(1):3-55.

Where can I buy?

If you would like more information or would like to purchase one of our monitors please select the country you are in on the right hand side. A full list of distributors in your area will be displayed.

PURCHASE