Quote Me

Medi-Gas MDM300

Advanced dew point hygrometer
Portable, fast and accurate moisture content analysis in the field

Medi-Gas MDM300

Summary

A simple-to-use monitor allowing more measurements per working hour than any other comparable hydrometer.

  • Fast, accurate and reliable measurement.
  • Lightweight and portable.
  • Bluetooth connection for wire-free data download.

Features

  • Menu-driven display.
  • User-friendly software and easy set-up.
  • Repeatedly fast measurements in under 10 minutes for T95 to -70°C dp.
  • Measurement parameters available at the touch of a button.
  • Long battery life (typically over 48 hours during use).
  • Best in its class – 1°C dp.
  • External 4-20mA device input for dew-point, temperature and pressure transmitters.
  • Bluetooth connection for trouble-free data download.
  • Built-in sampling system up to 300 bar, with external options up to 400 bar.
  • Lightweight – less than 1.5kg.
  • User friendly – just switch it on and use the menu to begin measuring.
  • Suitable for use when wearing protective gloves.
  • IP66/NEMA 4 rated, making it ideal for demanding outdoor applications.

Applications

  • Monitoring desiccant dryers for compressed air or plastic moulding equipment .
  • Moisture measurement in high-voltage switchgear quench gas.
  • Dew point in natural gas, process and pipelines.
  • Industrial gas quality.
  • Medical gas quality.
  • Dew-point measurement in metallurgical applications.
  • And many more….

How it works

The MDM300 hydrometer uses a highly-developed ceramic impedance sensor with state-of-the-art thin and thick film techniques.

The sensor adsorbs water vapour into a porous, non-conducting ‘sandwich’ between two conductive layers, built onto ceramic substrate base. The active sensor layer is less than one micron thick and the porous top conductor allows transmission of water vapour into the sensor of less than one nano-metre.

The resulting sensor responds rapidly to changes in moisture, both in detecting moisture and also when being dried. It is very rugged and gives 1°C dew-point accuracy, coupled with excellent long-term reliability and stability.

Sampling Kits

Medi-Gas Regulator Kit

This HTM02 compliant sampling kit has been specifically designed to safely extract gas from hospital outlets for use on our range of gas analyzers. It will regulate flow from gas outlets up to 10bar of pressure. The Medi-Gas Regulator kit comes with a set of gas specific probes and fitted with a quick connect system, to easily and quickly change between gas outlets without the need for tools. The kit is also oxygen clean and therefore can be used on oxygen outlets with out the risk of harm.